Call Us: (800) 701-0186

Tuttnauer EZ9P Automatic Autoclave with Printer (Rerfurbished)


Tuttnauer logo

Tuttnauer EZ9P Automatic Autoclave Steam Sterilizer w/Printer, Refurbished

Excellent Condition, Includes Pouch Rack

Tuttnauer EZ9 Fully Automatic Autoclave with Closed Door Drying – Refurbished in stock with 3 trays and 9″ x 18″ chamber 5-gallon volume capacity and a sterilization cycle data paper printer.

This tabletop autoclave is designed for the sterilization of wrapped and unwrapped instruments and related items found in dental, medical and veterinary clinics, first aid rooms, hospitals, laboratories, etc. A computerized control unit ensures a fully automatic sterilization cycle, control and monitoring of physical parameters.

Choose from three automatic programs, according to the material to be sterilized, and one auxiliary drying program. The autoclave is equipped with an air compressor that, during the drying stage draws air through a HEPA filter (0.2µm). It pushes the air through the heated chamber to remove moisture and facilitate the drying operation. Drying is performed with the door closed, preventing any contamination.

With the simplicity of one-touch design, all your sterilization and drying needs are fulfilled. The Tuttnauer EZ9 fully automatic autoclave will fill, sterilize, exhaust and dry at the touch of a button.


  • Closed door active drying system to maintain sterility and ensure efficient drying of packs and pouches.
  • High efficiency air pump circulates hot air after the sterilization cycle for fast drying. Hepa filter 0.2 µm air filter (replaceable).
  • Pre-loaded cycles for: unwrapped instruments, wrapped/packs, glassware and additional drying cycle.
  • All program parameters can be customized and stored.
  • Automatic shut off at the end of both the sterilization and dry cycles. Audible and visual cycle interruption alert.
  • Double safety locking device prevents door from opening while chamber is pressurized.
  • Low water sensor prevents activation of sterilization cycle when there is insufficient water in the chamber or reservoir.
  • Constant monitoring of temperature and pressure for sterilization.
  • Power outage recovery system.
  • Door design prevents steam from coming in contact with control panel.
  • Optional internal printer documents: date, time, temperature and pressure.
  • 510k clearance for dental handpieces.
  • Drain valve is located on the front, allowing for quick and easy draining of water reservoir.
  • International certifications
  • Sterilization Cycle Data Paper Printer

Note: Photo may look slightly different than actual unit

  • Chamber Size: 9″ x 18″
  • Chamber Volume: 5 Gal – 19 Liters
  • Overall Dimensions: 21.5″ D x 20″ W x 14.4″ H
  • Standard Cassettes Capacity: 2 full and 2 half
  • Tray Dimensions: 16.3″ x 6.7″ x 0.8″
  • No. of Trays: 3
  • Standard Unwrapped Cycle Time: Cold: 27 Min. Hot: 13 Min.
  • Voltage (V) – Frequency (Hz) 120 – 50/60
  • Power (W) – Current (A) 1400W – 12A
  • Autoclave Weight – Lbs (kg) 95 (43)

Materials:

  • Chamber is electro-polish and built of stainless steel 316 L.
  • Door is made of stainless steel CF8.
  • Trays are made of stainless steel 304.
  • Water reservoir is made of hard plastic material.
  • Door handle is made of hard plastic material, which is safe to touch and thermo-insulated.
  • Covers are made of aluminum sheet, coated with Epoxy paint.
Sterilization ProgramsTemp.Sterilization TimeDry Time
DescriptionIcon
Unwrapped Instruments & materials273° F3 min0
Wrapped Instruments & porous load273° F7 min30 min
Glassware250° F30 minslow exhaust:
15-20 min
Drying (continues the heat-assisted drying process)accessoryfor extra drying

Tuttnauer EZ9 Manual (PDF download)

Tuttnauer EZ9&10 Brochure (PDF download)

Shopping Cart